How to Calculate Readings for Compound Pressure Transducers

Oct. 11, 2013

Calculating the offset of the pressure transducer as it should read with no pressure (or the pressure on the bench) can be confusing without writing it down.

Different terms are used in the compressor industry when referring to pressure transducers that measure from negative to positive gauge pressure. Measuring both suction and discharge, a compound pressure transducer provides a linear output signal from full vacuum through zero pressure up to the full scale pressure.

A pressure transducer with a pressure range of -14.7 PSIG to 30 PSIG can be referred to as a compound pressure transducer, 30V30 (referring to -30" of mercury) transducer, or a vacuum to 30 PSIG sensor. While the terminology differs, it can refer to the same product. Calculating the offset of the pressure transducer as it should read with no pressure (or the pressure on the bench) can be confusing without writing it down. 

Here are a couple of examples to help walk you through the process for both 4-20mA output signals and voltage signals. The same principles apply to other pressures and output signals. We started with a pressure measurement range of -14.7 PSI to 15 PSI to help visualize how the calculation should work. While 0 PSI is close to the mid-point between the output, it is not exact; a few extra steps lead us to the right reading. 

4-20mA Example Model Number

AST4300AV0015P4N1000
AST4300 Non-Incendive Pressure Transmitter
1/4" NPT Male process connection
-14.7 PSI to 15 PSIG
4-20mA output signal
Conduit with 6' cable
316L stainless steel
First, you want to find out how many milliamps per PSI.  You will be measuring 14.7 + 15 = 29.7 PSI over the full scale pressure range
20mA – 4mA = 16 mA of span
16mA / 29.7 PSI = 0.5389 mA / PSI
Next, the "0" PSI pressure is going to be 14.7 PSI above the 4mA output, because the pressure transducer is scaled from -14.7 to full scale. 14.7 PSI  x 0.539  mA/PSI  = 7.923 mA
Finally ADD back in the 4mA for the zero offset (since the output starts @ 4mA and not 0 mA):
4 + 7.923 mA = 11.923 mA @ 0 PSI
4mA = -14.7 PSI
11.923 mA is the theoretical output of 0 PSI
20mA = 15 PSI full scale pressure
 

Voltage Example Model Number

AST4000KV0185P1F0143
AST4000 Industrial Pressure Transducer
SAE 4 female process connection with Schraeder Pin
-14.7 PSI to 185 PSI
0.5-4.5V ratiometric output signal
Packard Metripack 150
17-4 PH stainless steel
You will be measuring 14.7 + 185 = 199.7 PSI over the full scale pressure range
4.5V – 0.5V = 4V of span
4V / 199.7 PSI = 0.020V / PSI
Next, the "0" PSI pressure is going to be 14.7 PSI above the 4mA output, because the pressure transducer is scaled from -14.7 PSI to full scale.

14.7 PSI  x 0.020V / PSI  = 0.294V

Finally ADD back in the 0.5V for the zero offset (since the output starts @ 0.5V and not 0 V):
0.5 + 0.294441V = 0.794V @ 0 PSI
0.5V = -14.7 PSI
0.794V is the theoretical output of 0 PSI
4.5V  = 185 PSI full scale pressure
4-20mA Pressure Sensor Popular Compound Pressures

Pressure

Range

 Approximate Reading

at Zero Pressure

-14.7 to 15 PSI  11.92 mA
-14.7 to 30 PSI  8.93 mA
-14.7 to 50 PSI 7.48 mA
-14.7 to 100 PSI 6.05 mA
-14.7 to 185 PSI 5.18 mA
-14.7 to 300 PSI 4.75 mA
-14.7 to 500 PSI 4.46 mA

0.5-4.5V Ratiometric Pressure Sensor Popular Compound Pressures

Pressure

Range

Approximate Reading

at Zero Pressure

-14.7 to 15 PSI 2.48V
-14.7 to 30 PSI  1.73V
-14.7 to 50 PSI 1.52V
-14.7 to 100 PSI 1.01V
-14.7 to 185 PSI 0.80V
-14.7 to 300 PSI 0.69V
-14.7 to 500 PSI 0.61V
As the pressure increases, the amount of output signal for the compound range between -14.7 PSI and 0 PSI decreases. Once sensor error, such as zero offset is calculated, the ability to accurately measure the vacuum portion of the transducer diminishes. The pressures in the above charts are the most commonly calculated, although it is possible to calibrate higher pressures with compound pressure ranges.
Other pressure units, such as Bar and kg/cm2, are also available using the same methodology.  

This blog post was originally published on American Sensor Technologies' Pressure Transducers Blog at www.astsensors.com/blog/.

Sponsored Recommendations

Clean-in-Place (CIP) Solutions for Life Sciences Process Manufacturing

Learn how Emerson's measurement instrumentation can improve safety and reduce cross-contamination during CIP processes for life sciences process manufacturing.

Wireless Pressure Monitoring at Mining Flotation Cell

Eliminate operator rounds and improve flotation cell efficiency using reliable, wireless technology

Green hydrogen producer ensures quality of the network’s gas blend using a gas chromatograph

Case Study: Revolutionizing Green Hydrogen Blending with Precise Monitoring.

Overcome Measurement Challenges in Life Sciences

See how Emerson's best-in-class measurement instrumentation can help you overcome your toughest life sciences manufacturing challenges.